pg. 36 Rev E 1/14/15
The bag suspender should be checked every three to six months or if fiber values are higher than normal or
inconsistent.
Check the trays for melting.
The pictures shown are examples of extreme cases.
However, for proper operation you must replace trays
that show signs of melting or wear.
Check that the bottom tray is flat.
If the bottom tray is concave (see picture) the bag
suspender will catch in the vessel and melt.
Check the tip for excessive wear (see pictures).
Replace worn tips.
Troubleshooting
The ANKOM technology web site has the most current troubleshooting information. Therefore, if you have any
questions about the operation of your ANKOM
2000
Fiber Analyzer, please visit our web site at www.ankom.com.