Operator’s Manual
pg. 34 Rev E 1/14/15
If you see a leak
If you see a leak, follow steps 6 9 in the Initial Maintenance procedure above.
Replacing the Fuses
To replace the fuses in the ANKOM
2000
Fiber Analyzer, follow the procedure detailed below.
1.
Turn off the instrument power and unplug the
Power Cord from the outlet.
2.
Using a flat blade screwdriver, twist the slot on
the fuse holders counterclockwise ¼ turn to
open.
3.
Check both fuses.
4.
Pull the fuse from the grey fuse cap and
replace as needed.
(120V 15 amp / 220V 10 amp)
Cleaning the Fiber Optic Level Sensor
Using a cotton swab with alcohol, wipe the tip of the
level sensor at least once per month if you use the
instrument daily.
Level Sensor
Fuses (2)