Operator’s Manual
Rev F 12/08/14 pg. 25
Check Temperature Sensor
To check the Temperature Sensor, follow the procedure below.
1. Access the Diagnostics screen (see "Access Diagnostics Screen"
section for details).
2. With the Diagnostics screen displayed, press the ENTER button on
the Keypad. The screen to the right will be displayed.
3. Press the Down Arrow button (v) on the Keypad. The screen to
the right will be displayed.
4. Press the ENTER button on the Keypad. The screen to the right
will be displayed showing the current Temperature Sensor
reading.
When the instrument has not been run for more than two hours,
the temperature should read the ambient temperature of the room.
If the temperature reading does not show ambient temperature
when the instrument is cool, contact ANKOM Technology.
5. Press the STOP button on the Keypad to go back to the Analog
screen (shown to the right).
6. Press the STOP button on the Keypad to go back to the main
Diagnostics screen (shown to the right).
7. Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more
times to return to your solvent screen (example shown to the
right).
PET Ether
v <ENTER>
Diagnostics
^ <ENTER>
Analog <ENTER>
Exit <STOP> ^ v
Temp 20°C
Analog <ENTER>
Exit <STOP> ^ v
Outputs <ENTER>
Exit <STOP> v
Diagnostics